6 Troubleshooting
Sirona Dental Systems GmbH
6.8 Error analysis of X-RAY control signal path
Service Manual GALILEOS
Error messages
with and without installed remote control
Error code
Description
Actions required
see
E3 13 40
Short circuit in signal path between
board DX11 and release button A2
during power-on.
● Replace cable L117 or L108.
● Replace board DX1.
● Replace board DX11.
E6 13 41
Release signal missing on board
DX11 at start of exposure.
● Replace cable L117 or L108.
● Replace board DX1.
● Replace board DX11.
E3 13 42
Short circuit in signal path between
board DX11 and release button A2
during operation of the unit.
● Replace cable L117 or L108.
● Replace board DX1.
● Replace board DX11.