61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
117
Sirona Dental Systems GmbH
5
Messages
Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
E6 71 12
E6 71 20
E6 71 21
Error code
Description
Actions required
see
E7 71 12
Unit is not ready for operation
● Check the CAN bus.
This error is a sequential fault.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
Error code
Description
Actions required
see
E6 71 20
Contact to DX11 interrupted during
operation.
● Note error message on remote control
(DX42) and check log memory (via
extended details).
● Check the CAN bus.
● Check cable L9, replace if necessary.
S. [ → 125],
S. [ → 128],
S. [ → 141],
S. [ → 362]
Error code
Description
Actions required
see
E7 71 21
No CAN bus connection. DX11 does
not start.
Occurs in the start screen after
power-on.
● Start the detail query via SiXABCon.
S.
If board DX11 responds ...
● Check the signal path to DX71, repair or
replace cables/connectors if necessary.
● Replace board DX1.
If board DX11 does not respond ...
● Replace board DX11.