61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
259
Sirona Dental Systems GmbH
8
Service
Service Manual GALILEOS
8.11
S017: Configuration service
båÖäáëÜ
8.11.5
S017: Test step 6
S017: Test step 6
Enable/disable the remote control
* Factory setting
S017_6 GALILEOS
1. Call service routine S017.6 [ → 213].
ª
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
Selection field
Code
Function
1
00
Remote control
disabled*
01
Remote control
enabled