Messages
Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
E6 06 60
E6 06 65
E6 06 66
E6 06 67
E6 06 68 GALILEOS
Error code
Description
Actions required
see
E6 06 60
TDI signal from board DX11 to
board DX6 is disturbed.
TDI = Signal to start synchronized
readout sequence and to prepare
the next exposure
● Replace cable L15.
● Replace board DX1.
● Replace the tube assembly.
S. [ → 362],
S. [ → 342],
S. [ → 310]
Error code
Description
Actions required
see
E6 06 65
Tube current or tube voltage is too
high in standby mode.
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 66
Impermissible tube type.
● Check tube type of tube assembly
using extended detail query or service
routine S005.1, replace tube assembly
if necessary.
Error code
Description
Actions required
see
E6 06 67
Light guide input TDI is active during
switch-on.
TDI = Signal to start synchronized
readout sequence and to prepare
the next exposure
● Check TDI signal:
1. Switch off the unit.
2. Disconnect cable L15 at board DX11.
3. Switch unit on.
4. Perform visual check at socket J5:
● If light is visible: Replace board DX11.
● If no light is visible: Replace the tube
assembly.
Error code
Description
Actions required
see
E6 06 68
Tube assembly output after
exposure does not match the
expected value.
● Replace the tube assembly.