61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
223
Sirona Dental Systems GmbH
8
Service
Service Manual GALILEOS
8.5
S005: General X-ray tube assembly service
båÖäáëÜ
8.5
S005: General X-ray tube assembly service
S005: General X-ray tube assembly service
* SR=service routine, ** SHZ=security access
8.5.1
S005: Test step 1
S005: Test step 1
Read or select the X-ray tube assembly type
Tube type GALILEOS
➢ Call service routine S005.1 [ → 213].
ª
Selection field 1 shows the indicator number of the installed tube
type.
S005.1 GALILEOS
SR*
SA**
Function
S005
General X-ray tube assembly service
S005.1
No
If tube type is valid:
Read out tube type
Yes
If tube type is not valid:
Select tube type
S005.4
No
Fan test
S005.5
No
Temperature sensor test, single tank
S005.8
Yes
Automatic adjustment of pulse preheating
IMPORTANT
Only CB tube D151 R (indicator number 03) is permissible for use with
the GALILEOS volume tomography unit.
IMPORTANT
The X-ray tube assembly automatically queries the tube type
information. If no defined value is saved, the security access will be
shown instead of the tube type.