7 Adjusting/calibrating the unit
Sirona Dental Systems GmbH
7.2 Checking the mechanical system adjustment
Service Manual GALILEOS
7.1.6.3
Constancy test phantom
Constancy test phantom
The constancy test phantom is inserted in the bite block holder of the unit
for for the constancy and acceptance tests as well as for the check of the
mechanical unit adjustment [ → 166].
7.2
Checking the mechanical system adjustment
Checking the mechanical system adjustment
1. Insert the constancy test phantom in the bite block holder of the unit.
2. Measure distances B1, B2and B3 between the tube assembly
housing and measuring point M on the constancy test phantom
(positions 1, 2 and 3) using the steel tape measure from the service
set.
3. Then calculate the ideal distance between the tube assembly and
measuring point M as follows: (B1+B2)/2 = ideal distance
4. Distances B1, B2 and B3 must not deviate more than ± 2 mm from
the calculated ideal distance. With deviations exceeding ± 2 mm, the
unit must be adjusted mechanically via the position of the ring
M
B1
P1
M
B2
P2
M
B2
P3
M