8.5 S005: General X-ray tube assembly service
Service Manual GALILEOS
If the tube type is invalid ...
1. Confirm the security access [ → 215].
ª
Selection field 1 shows the indicator number of the (invalid) tube
type detected by the X-ray tube assembly.
2. Use the arrow keys in selection field 1 to select the code for the tube
type (see table).
ª
The Memory key (B) (Easypad) or the LED above the Memory
key (B) (Multipad) lights up.
3. Save the selected parameter [ → 219].
4. Exit the service routine [ → 220].
After calling the service routine S005.1 again, the security access is no
longer requested and the stored tube type is displayed.