61 25 665 D3437
110
D3437
.
076.01.15
.
02 06.2012
5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
E6 41 23
E6 41 24
E6 41 25
Error code
Description
Actions required
see
E6 41 23
Hardware fault at controller input on
board DX41.
Board DX41 detects a wrong signal
level of the hardware signal for
radiation release.
● see section Error analysis of X-RAY
control signal path: up to unit serial
number 3199 (with board DX41) [ → 142].
Error code
Description
Actions required
see
E6 41 24
Short circuit in radiation release
signal path between board DX42
and board DX41 (cable L17).
The release signal was detected on
boards DX11 and DX41 but not on
board DX42.
● see section Error analysis of X-RAY
control signal path: up to unit serial
number 3199 (with board DX41) [ → 142].
Error code
Description
Actions required
see
E6 41 25
X-RAY hardware signal present,
software signal not present.
● Check cable L17, replace if necessary.
● Check board DX42, replace if necessary. S. [ → 134],
● Check board DX41, replace if necessary. S. [ → 134],