Service
Service Manual GALILEOS
8.4
S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time
båÖäáëÜ
8.4
S002: Radiation without rotary movement,
selectable kV/mA level and maximum radiation
time
S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time
* SR=service routine, ** SHZ=security access
8.4.1
S002: Test step 5
S002: Test step 5
Long-term exposure with fixed radiation intervals from any position
* Factory setting
SR*
SHZ**
Function
S002
X-ray beam test
S002.5
Yes
Long-term exposure with fixed radiation intervals from any position
WARNING
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
Selection field
Parameters
Range of values
1
kV/mA level
60 kV/8 mA –
85 kV / 7 mA*
2
Radiation time
0.1 s – 5.0 s*