Connections with POLYDOROS SX 65

4 - 17

Siemens AG

 

RXL2-120.812.01

Page 17 of 24

LITHOSTAR Multiline

Medical Solutions

Rev. 02

10.04

CS PS 24

System

 

Connect the protective conductor cable to H1.

 (2/Fig. 12).

Thread the (primary) leads H1.U1 and H1.V1 of the inverter module per the designation 
onto H1 (3/Fig. 13).

Route the cable connected to D220.X41 (4/Fig. 13) to board D160 and connect it to plug 
X44.

Plug the cable X1 from board D100 to connector X1 on board D220 (5/Fig. 12) and 
strain-relieve it on the cover (shielding underneath the strain relief (6/Fig. 12)).

Reinstall the guard with the 

 symbols (1/Fig. 12).

Check the nuts on the protective conductor studs (7/Fig. 13) and tighten them, if neces-
sary.

Summary of Contents for LITHOSTAR Multiline

Page 1: ...tion transmission or use of this document or its contents is not permitted without express written authority Offenders will be liable for damages All rights including rights created by patent grant or registration of a utility model _or_ design _are_ reserved SP System for Subcontractors and Siemens Technicians Installation Instructions ...

Page 2: ...levels Disclaimer The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho rized by Siemens or one of its affi...

Page 3: ...ing the Cooling Unit 2 4 Setting Up the Unit 2 4 Installing the Covers 2 7 Installing the Components for the System Controller 2 9 Setting Up the Console Table 2 9 Positioning Connecting the Monitors 2 9 Positioning the Keyboard and the PC Mouse 2 9 Connecting the Footswitch Tableside Control Console Shock Wave Trigger Mechanism 2 9 Back panel of the PC 2 10 Positioning Connecting the Operating Sy...

Page 4: ...A12 electronics cabinet 4 2 Overview of cable run in the power cabinet 4 3 Connections harness locations in the generator cabinet Part A 4 4 Connections of the options and their cable positions in the A12 electronics cabinet and generator AC 4 6 Connections and cable locations in the A12 electronics cabinet and in the control console 4 8 Connections cable locations in the A13 shock wave cabinet 4 ...

Page 5: ...in the Planning Guide PG RXL2 120 021 01 The hardware for mounting the unit to the floor Liebig expansion bolts mounting plate must be installed per Installation Instructions RXL2 120 038 01 The condition of the access route within the building as well as the condition of the room in which the unit is to be installed must be in compliance with the project drawings and the data listed under the Gen...

Page 6: ...yellow red enamel are shipping brackets which may be removed only in accordance with the Installation Instructions Test Equipment and Tools Service Tools Cataloque 1 Installation auxiliary devices 1 Materials 1 Tool kit for Installation engineers Torque wrench 20 to 100 Nm Digital Multimeter Transport carriage roller set 11 53 654 G5338 The roller set must be ordered if the system is not installed...

Page 7: ... Patient monitor Option SC 3rd Monitor Option Polydoros LX AC A12 A13 HPE charger N12 Shock wave controller Cooling unit VIDEO MED DIM Image storage N11 System controller M16 Power supply Water prepara tion PC Monitor Monitor I II Multispot or DICOM Connect Options Operating system unit Ultrasound unit Option Diamentor Option DICOM Monitor Option ...

Page 8: ... Polydoros SX AC A12 A13 HPE charger N12 Shock wave controller Cooling unit VIDEO MED DIM Image storage N11 System controller M16 Power supply Water prepara tion Patient monitor Option SC 3rd Monitor Option PC Monitor Monitor I II Multispot or DICOM Connect Options Operating system unit Ultrasound unit Option Diamentor Option DICOM Monitor Option ...

Page 9: ...lation work 2 Check the condition of the table installation plate floor plate Check the cable feed throughs for cabinets system control console and the unit for usability Options Monitor trolley and or Patient monitor 2 Check the cable feed throughs for the Patient monitor for usability Make sure that there is sufficient space for a Patient monitor p p p p ...

Page 10: ...ed location using the supplied template Drill the holes and insert the Liebig anchors that are included in the shipment Fig 2 Remove all transport safety brackets from the generator Take the generator off the pallet If necessary knock out the cable inlet in the generator cabinet for interconnections with the adjacent cabinet Install the edge protection strip service package Mount the generator in ...

Page 11: ...binet A13 If necessary turn it around Install the sidewalls of the cabinets Within the scope of validity of the UL regulations e g USA the metal case should not be removed due to fire protection Install the enclosed screw 1 Fig 3 with nut 2 Fig 3 to a cabinet to connect the cabinet Then move the second cabinet to its position and tighten with nut 4 Fig 3 and flat washer 3 Fig 3 Move the cabinets t...

Page 12: ...e Unit 2 Place the four transport rollers 2 Fig 4 onto the transport carriage on the yellow shipping frame at c Fig 5 and bolt them in place Remove the shipping pallet from the unit Lift the unit with the help of the spindles and pull the transportation pallets out to the side Depending on the dimensions of the transport route it may be necessary to reposition the rollers on the carriage Pos a or ...

Page 13: ...red Uncoil the four cable harnesses that are permanently connected to the unit and lay them out Fig 7 Guide the free ends of the cable harnesses into the opening in the mounting plate and into the corresponding cabinets via the cable conduits The cable harnesses are color coded blue for the shock wave cabinet green for the control cabinet yellow for the generator The correspondingly color coded ta...

Page 14: ...he W drive Measure the resistance to ensure that the unit is electrically isolated on the mounting plate Carefully feed through the cable harnesses Store any excess cable lengths Roll them up without kinks Do not place any excess cable or cable loops in the unit frame since they may come into contact with the moving energy chain and become damaged Do not block the movement of the energy chain Remo...

Page 15: ... plastic bags Secure the two lateral strips on the unit base frame with three paneling screws for each Fig 10 Install the rear metal roll down sash in the unit base frame on the left and the right Fig 11 for this the mounting brackets must be installed on the inside of the frame Attach the cover K Fig 11 for the cable routing using two paneling screws Before installing the covers connect each meta...

Page 16: ...sound option is being installed Insert the two part base frame cover and mount it in place using four paneling screws for each part Fig 12 Install the kick plates on both table legs and fasten them together Fig 13 Attach the paneling to the underside of the shock head support A Fig 14 In the front section of the base frame B Fig 14 use black plastic plugs to seal the two holes for mounting the tra...

Page 17: ... console table as shown in Fig 15 and connect them per their designations Positioning the Keyboard and the PC Mouse 2 Place the keyboard and the PC mouse on the control console table Connect the cable to the keyboard Connecting the Footswitch Tableside Control Console Shock Wave Trigger Mechanism 2 Connect the tableside control without the holder optional B Fig 16 and the C Fig 16 shock wave relea...

Page 18: ...e to the operating system unit Insert the slide out shelf Pay attention to the cable routing in the console Cover the cable outlet 2 In areas where UL regulations apply e g USA you must cover the cable outlet with the extra aluminum sheet located in the operating console as fire protection Positioning Connecting DICOM Connect Option 2 Perform installation and startup according to the manual RXL2 1...

Page 19: ...the Diamentors Option 2 Plug the power cable into the multiple socket outlet of the console desk slide out shelf Lead the cable up through the free opening Connect the cable routed from the measuring chamber of the Diamentor via the adapter cable supplied Connect the cable routed from power supply unit at the Diamentor Position the Diamentor on the console desk The two Allen screws in the base of ...

Page 20: ...D115 X2 under the clamp Reinstall the TV camera housing 1 Fig 18 but do not secure it Connect the protective conductor to the white cover 2 Fig 18 Place the white cover 2 Fig 18 over the TV camera and secure it together with the housing Use the white plastic plugs to cover the holes Installing the TV camera on the 33 I I 2 The camera has been pre installed at the factory Checking the temperature o...

Page 21: ...f the blades close to the focus with the template Attach the collimator to the tube unit Plug in connector X1 on the collimator Secure the wire to the panel using strain relief clamps 1 Fig 19 Connect the plug Installing the Diamentor Option 2 Remove the cover from the collimator Insert the measuring chamber and lead the cable out through the opening 1 Fig 20 Reattach the cover to the collimator P...

Page 22: ...o the holder of the monitor trolley The monitor should lock into position Tighten the two rear screws Connect the cables according to the labels Clamp the shielding of the BAS lead under the clamp of the VIDEOIN cable Establish the protective conductor connections Reinstall the monitor housing Refer also to installation instructions RX53 020 031 06 System with Polydoros LX chapter 3 System with Po...

Page 23: ...arly in the cabinets Do not store excess cable lengths in the cabinets EMC reasons Be careful that the cable harnesses do not get caught in the swing out frames Water hoses 5 and 8 may only be shortened by max 5 m X ray generator AC yellow Shock wave cabinet A13 blue Electronics cabinet A12 green Do not open the connector housing on the cable between the camera and the VIDEOMED DI M D8 X1 under an...

Page 24: ...n in the adjacent Fig 2 and connect them according to table 1 and 3 The numbers or letters e g are located in the cable lists on the following pages Prior to routing the system cables remove the EMC shielding plate at N11 D11 right cable harness green from control console centre cable harness green from unit K3 K4 left cable harness green from unit K8 K9 4 VIDEOMED DIM Mounting rail for cable harn...

Page 25: ...connect them as indicated in Tables 1 and 2 Multiline power line connections Connect the power harness for the Multiline in the M16 at terminal strip K30 Fig 3 see also the POLYDOROS LX Wiring Diagram X2169 11 K30 L1 L2 L3 is switched over the GS breaker with generator ON K30 N and K30 terminal pins Fig 3 Strain relieve the power harness shielding under the clamp Fig 3 X5 X4 X3 D110 D115 D100 D200...

Page 26: ...e using cable ties AC Generator A12 Electronics Cabinet A13 Shock Wave Cabinet Connections Type cable connection Function Generator AC A12 Cabinet A13 Cabinet ok 4 x 4 mm2 Gnd shielded MULTILINE power supply M16 k30 L1 L2 L3 N M16 X1 N F7 F8 F9 1 x 6 mm2 Protective con ductor M16 k30 M16 9 pole Canon XCS D200 X3 VDIM D5 X13 9 pole Canon Act value interface D100 X200 N11 D6 X8 3 x 1 5 mm2 shielded ...

Page 27: ... x 16 mm2 Cable harness in the middle Protective conductor Device PE M16 PE 3 x 1 5 mm2 shielded cable harness in the middle Power connection Z96 Device X11 1 2 3 M16 X2 L1 N PE 4 x 2 5 mm2 shielded cable harness in the middle Voltage supply of the output stages Device K1 M16 D21 X14 L1 2 3 x 1 5 mm2 shielded cable on the right Power connection control console Socket control con sole M16 X2 L2 N P...

Page 28: ...n Position the patient monitor in the location provided per the project planning guide e g ceiling support Plug the cable into the SYNC connector of the patient monitor Connect the patient monitor to its own power supply i e one that is independent of the LITHOSTAR Multiline Do not establish any protective conductor connections to the LITHOSTAR 9 CAUTION VIDEOMED DIM Rail for cable harness Cable h...

Page 29: ...n the AC Generator ok 3 x 1 5 mm2 Line voltage 3rd Monitor A12 M16 X2 L2 N PE 1 x 4 mm2 Protective conduc tor Monitor trolley A12 M16 Coax cable Video Monitor trolley A12 Monitor 3in Radiation ON dis play Monitor trolley AC D160 X6 see generator wiring diagram X2169 11 K14 ECG control Patient monitor SYNC A12 N11 D11 k14 Tab 4 The monitor cable contains more cable than necessary for connecting the...

Page 30: ...C shielding on the Videomed Route the camera connector D8 X1similar to the video cable underneath the rack and connect it to the backplane as shown in Table 5 Remove board D1 for this purpose Attach the cable with a cable clip to the rack above the shield Attach the EMC shielding on the Videomed Connect the XCS cables X11 and X12 on the D5 board 8 8 7 6 9 4 4 5 10 A12 Mounting rail for cable harne...

Page 31: ... console Tab 5 3 Remove EMC shielding on the D6 board Insert cable Reattach the EMC shielding and make sure that the cable is in the recess of the EMC cover 4 After connecting D8 X1 install the EMI shielding plate refer also to VIDEOMED connection in Chapter 3 page 8 Unit A12 Electronics Cabinet Connections center Cable Harness green Type of cable connector Function In Unit In Cabinet ok 4 pole co...

Page 32: ...le Canon Controller Slide out shelf N11 D11 K11 9 pole Canon 4 channel ECG monitor 4 channel ECG monitor N11 D11 K103A Coax video cable Video input Monitor 1in 1 Video monitor Monitor 1in MEMOSKOP X11 8 Coax video cable Video input Monitor 2in 2 Video monitor Monitor 2in MEMOSKOP X12 8 Coax video cable Video connection Monitor 1 Monitor 3 3 Video monitor Monitor 1out Monitor 3in 9 7 o to 3rd monit...

Page 33: ...n in Table 8 Secure the cables leading to N12 with cable ties on the right side of position 50 52 Make sure the conductors to D19 and D91 have sufficiently long For connecting the cables to the charger you have to tilt M12 toward the front Use cable ties to secure the cables on the right side of position 67 68 Connect the hoses for the water connections at the input output point in the shock wave ...

Page 34: ...ock head sensors X11 Shock wave head N12 K101 9 pole Canon Coding Shock wave head N12 K102 Special K26 charger lead X1 Power unit Charger 6 1 x 4 mm2 Protective conductor Power unit PE Charger 6 Type of hose Water Cooling system intake Shock wave head 5 Exchanger Water Cooling system return Shock wave head 8 Exchanger Tab 8 5 Allow enough cable length so that the boards can be unplugged with the c...

Page 35: ...of the cabinet near the floor Fig 8 Rotating anode cable connection Route the rotating anode cable internally required length approx 1 m up along the right side in the frame of the generator cabinet to the anode starter Fig 8 Connect the braided shielding external shielding of the end of the lead to the right side of the starter unit under one of the clamps Fig or 1 Fig 9 Connect the red cable inn...

Page 36: ...he high voltage receptacles approximately 1 cm if necessary refill with oil part no 44 10 023 RV090 1 liter Insert the high voltage plug without silicone disk but with silicone gasket into the correct receptacle of H1 Workstation 1 Generator Unit Connections Cable Harness yellow Type of Cable Function Generator System ok Stator cable 3 x 0 75 mm2 Rotating anode drive Anode starter K31 R2 R4 K32 R2...

Page 37: ...e cables 3 Fig 11 Carefully remove the plastic guard plate for board D220 with symbols 1 Fig 11 Connect the protective conductor strap to the H1 2 Fig 11 Attach the primary leads H1 U1 and H1 V1 of the inverter module per the designation onto H1 3 Fig 12 Route the cable connected to D220 X41 4 Fig 11 12 down to the D160 board and connect it to X44 Clamp the braided shield of the cable under the cl...

Page 38: ... clamps 180 both curva tures opposite to each other Insulate the harness terminals that have been installed using heatshrink tubing Remove the cover over the knife breakers F1 F2 and F3 Fig 13 at the M16 of the generator Connect the power line leads L1 L2 L3 N PE to the fuses F1 F2 F3 connection Fig 13 and to terminal strip K20 N and Strain relieve the power cable on the bottom plate of POLYDOROS ...

Page 39: ...quired it must be connected in series with the oil pressure switch Have the door contact activated e g by the examination room door External radiation indicator lamp Remove the corresponding jumper and connect the function according to wiring diagram X2169 11 If necessary pull the left part of the D160 X6 plug out from the side Generator Periphery Type of Cable Function Generator ok Power line cab...

Page 40: ...his is possible only after moving the system Route the fiber optic and video cables to the control cabinet Refer to the installation instructions for fiber optic cables module RX0 000 038 02 Route the video cable into the control cabinet not used Connect the ultrasound tableside control to SONOLINE Connect the ultrasound probe and place the probe in its holder Move board D6 slightly away from N11 ...

Page 41: ...L2 120 812 01 Page 19 of 20 LITHOSTAR Multiline Medical Solutions Rev 02 10 04 CS PS 24 System Fig 15 Power cable connector Clamp for power cable A D31 Protective conductor Clamp for corrugated hose Corrugated hose from SONOLINE Cable outlet to cabinets ...

Page 42: ...plug Sonoline 3 Remove the existing screw at point 1 Fig 17 Screw in the mounting screw according to Fig 17 The clamp 2 Fig 17 prevents loosening the plug connection without tools Tighten the counternut 1 Fig 17 Fig 16 CE approval is valid for the entire system only If the ultrasound unit is removed then this does not retain the required CE appro val 1 2 NOTE ...

Page 43: ...y in the cabinets Do not store excess cable lengths in the cabinets EMC reasons Be careful that the cable harnesses do not get caught in the swing out frames Water hoses 5 and 8 may only be shortened by max 5 m X ray generator AC yellow Shock wave cabinet A13 blue Electronics cabinet A12 green Do not open the connector housing on the cable between the camera and the VIDEOMED DI M D8 X1 under any c...

Page 44: ...n in the adjacent Fig 2 and connect them according to tables 1 and 3 The numbers or letters e g are located in the cable lists on the following pages Prior to routing the system cables remove the EMC shielding plate at N11 D11 right cable harness green from operat console center cable harness green from unit K3 K4 left cable harness green from unit K8 K9 4 VIDEOMED DIM Mounting rail for cable harn...

Page 45: ... 812 01 Page 3 of 24 LITHOSTAR Multiline Medical Solutions Rev 02 10 04 CS PS 24 System Overview of cable run in the power cabinet 4 Fig 3 M9 DIA Route the cable in the power cabinet from the left or right only E J B D H 8 T2 M16 L X3 17 16 A L C K5 K32 K31 ...

Page 46: ...1 T1 L2 T2 L3 T3 N K30 N PE Bolt on M16 AC Generator A12 Electronics Cabinet A13 Shock Wave Cabinet Connections Type cable connection Function Generator AC A12 Cabinet A13 Cabinet ok 4 x 4 mm2 Gnd shielded MULTILINE power supply M16 L1 L2 L3 N M16 X1 N F7 F8 F9 1 x 6 mm2 Protective con ductor M16 k30 M16 9 pole Canon XCS D 320 X3 VDIM D5 X13 9 pole Canon Act value interface D100 X200 N11 D6 X8 3 x...

Page 47: ...le 2 Fig 3 Strain relieve the harnesses on the side using cable ties Power and voltage supply connection protective ground wire connection in A12 Type of cable connection Function Device operating console Cabinet A12 ok 1 x 16 mm2 Cable harness in the middle Protective conductor device PE M16 PE 3 x 1 5 mm2 shielded cable harness in the middle Power connection Z96 device X11 1 2 3 M16 X2 L1 N PE 4...

Page 48: ...osition the patient monitor in the location provided per the project planning guide e g ceiling support Plug the cable into the SYNC connector of the patient monitor Connect the patient monitor to its own power supply i e one that is independent of the LITHOSTAR Multiline Do not make any direct ground wire connections to the LITHOSTAR 9 CAUTION VIDEOMED DIM Mounting rail for cable harness Cable ha...

Page 49: ...AC Generator ok 3 x 1 5 mm2 Power supply voltage 3rd Monitor A12 M16 X2 L2 N PE 1 x 4 mm2 Protective conduc tor Monitor trolley A12 M16 Coax cable Video Monitor trolley A12 Monitor 3in Radiation display Monitor trolley AC D160 X61 refer to generator wiring diagram X2206 11 K14 ECG control Patient monitor SYNC A12 N11 D11 k14 Tab 4 The monitor cable contains more cable than necessary for connecting...

Page 50: ... the EMC shielding on the Videomed Route the camera connector D8 X1similar to the video cable underneath the rack and connect it to the backplane as shown in Table 5 Remove board D1 for this purpose Attach the cable with a cable clip to the rack above the shield Attach the EMC shielding Reinstall the XCS cables X11 and X12 on the D5 board 8 8 7 6 9 4 4 5 10 A12 Mounting rail for cable harness Cabl...

Page 51: ...Remove the EMC shielding plate on the D6 board Insert the cable Reattach the EMC shielding plate and make sure that the cable is in the recess of the EMC cover 4 After connecting the D8 X1 reinstall the EMI shielding plate Refer also to VIDEOMED connection in chapter 4 page 8 Unit A12 Electronics Cabinet Connections center Cable Harness green Type of cable connector Function In Unit In Cabinet ok ...

Page 52: ...ntroller Slide out shelf N11 D11 K11 9 pole Canon 4 channel ECG monitor 4 channel ECG monitor N11 D11 K103A Coax video cable Video input Monitor 1in 1st Video moni tor Monitor 1in MEMOSKOP X11 8 Coax video cable Video input Monitor 2in 2nd Video moni tor Monitor 2in MEMOSKOP X12 8 Coax video cable Video connection Monitor 1 Monitor 3 3rd Video moni tor Monitor 1out Monitor 3in 9 7 o to 3rd monitor...

Page 53: ...n in Table 8 Secure the cables leading to N12 with cable ties on the right side of position 50 52 Make sure the conductors to D19 and D91 are sufficiently long For connecting the cables to the charger you have to tilt M12 toward the front Use cable ties to secure the cables on the right side of position 67 68 Connect the hoses for the water connections at the input output point in the shock wave c...

Page 54: ...on Shock head sensors X11 Shock wave head N12 K101 9 pole Canon Coding Shock wave head N12 K102 Special K26 charger lead X1 Power unit Charger 6 1 x 4 mm2 Protective conductor Power unit PE Charger 6 Type of hose Water Cooling system intake Shock wave head 5 Exchanger Water Cooling system return Shock wave head 8 Exchanger Tab 8 5 Allow enough cable length so that the boards can be unplugged with ...

Page 55: ...0 812 01 Page 13 of 24 LITHOSTAR Multiline Medical Solutions Rev 02 10 04 CS PS 24 System Overview of cable run in the power cabinet 4 Fig 8 M9 DIA Route the cable in the power cabinet from the left or right only E I B D H T2 M16 F X3 17 16 A L C K31 K5 K32 ...

Page 56: ... the shielding of the zipper hose to the bolt in the M16 Fig 8 Strain relieve the zipper hose Rotating anode cable connection Guide the rotating anode cable internally required length approx 1 m up along the right side in the frame of the generator cabinet to the anode starter Fig 8 Connect the braided shielding external shielding of the end of the lead to the right side of the anode starter under...

Page 57: ...roximately 1 cm if necessary refill with oil part no 44 10 023 RV090 1 liter Insert the high voltage plug without silicone disk but with silicone gasket into the correct receptacle of H1 Workstation 1 Slide the H1 high voltage transformer into the generator cabinet Generator Unit Connections Yellow Cable Harness Type of Cable Function Generator System ok Stator cable 3 x 0 75 mm2 Rotating anode dr...

Page 58: ...tion of the transformer mount Fig 10 lift it and snap in the bayonet connectors 1 Fig 10 of the mount Use cable ties to strain relieve the cables 8 Fig 13 Carefully remove the plastic guard plate for board D220 with symbols 1 Fig 12 Fig 11 Fig 12 Fig 13 There must be a ground connection between the cover and ground 2 1 6 5 U1 U2 V1 V2 U1 U2 V1 V2 3 4 7 8 NOTICE ...

Page 59: ...1 of the inverter module per the designation onto H1 3 Fig 13 Route the cable connected to D220 X41 4 Fig 13 to board D160 and connect it to plug X44 Plug the cable X1 from board D100 to connector X1 on board D220 5 Fig 12 and strain relieve it on the cover shielding underneath the strain relief 6 Fig 12 Reinstall the guard with the symbols 1 Fig 12 Check the nuts on the protective conductor studs...

Page 60: ...G System Rev 02 10 04 CS PS 24 Medical Solutions 4 18 Connections with POLYDOROS SX 65 Overview of cable run in the power cabinet 4 Fig 14 M9 DIA Route the cable in the power cabinet from the left or right only E J B D H 8 T2 11 M16 L X3 17 16 A L C K5 K32 K31 ...

Page 61: ...t to the conductors of the power cable If the power cable diameter is small enough turn the clamps 180 both curvatures opposite to each other Apply end splices to the power cable Connect power leads L1 L2 L3 N PE connection Fig 13 to terminal strip K20 Strain relieve the power cable Existing shielding should not be connected Generator Periphery Type of Cable Function Generator ok Power line cable ...

Page 62: ...ces and indicators can be connected in the POLYDOROS SX to D160 X61 to 64 1 Fig 15 Connection of the door contact for blocking radiation After removing the associated jumper connect the door contact Connection of the indicator lamps for tube assembly selection Attach an indicator lamp in the vicinity of the tube assembly 24 V max 5 W so that it is clearly visible from the control console Connect t...

Page 63: ...p for AP1 Connect the temperature monitoring devices and oil pressure switches Room lighting control A contactor for the room lighting control can be connected to board D160 To program the room lighting control refer to the Start up instructions System configuration RXL2 120 034 Site Adjustments in the section Room light as required Connect contactor connection to Indication with ZB Indication wit...

Page 64: ...he control cable to D31 X28 Fig 16 This is possible only after moving the system Route the fiber optic and video cable to the control cabinet Adhere to the installation instructions for fiber optic cables module RX0 000 038 02 Route the video cable into the control cabinet not used Connect the ultrasound tableside control to SONOLINE Connect the ultrasound probe and place the probe in its holder M...

Page 65: ...L2 120 812 01 Page 23 of 24 LITHOSTAR Multiline Medical Solutions Rev 02 10 04 CS PS 24 System Fig 16 Power cable connector Clamp for power cable A D31 Protective conductor Clamp for corrugated hose Corrugated hose from SONOLINE Cable outlet to cabinets ...

Page 66: ...plug Sonoline 4 Remove the existing screw at point 1 Fig 17 Screw in the mounting screw according to Fig 17 The clamp 2 Fig 17 prevents loosening the plug connection without tools Tighten the counternut 1 Fig 17 Fig 17 CE approval is valid for the entire system only If the ultrasound unit is removed then this does not retain the required CE appro val 1 2 NOTE ...

Page 67: ... this should be done after start up has been completed as described in the Start up Instructions RXL2 120 034 Install the two large plastic panel sections on the side of the unit and mount them in place Fig 1 to do this temporarily loosen the two screws on the beveled surfaces of the rear cover panel Install the X ray tube housing and attach them together Fig 2 Secure the EMC metal shielding with ...

Page 68: ... panels Seal all seams between the system and the floor with silicone suitable for wet cells Fig 4 and Fig 5 Complete the installation protocol in register 3 of the log book Adapting the Other Accessories 5 Perform a test installation of all accessories that have been included in the shipment in accordance with the Operating Instructions System Insulating plate Floor Silicone Silicone Mounting pla...

Page 69: ...L2 120 812 01 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev 02 10 04 CS PS 24 System Changes to previous version 6 Chap 0 Cover Contents Revision newly generated Chap 4 24 Page new added Chap 3 20 Page new added ...

Page 70: ...LITHOSTAR Multiline RXL2 120 812 01 Page 2 of 2 Siemens AG System Rev 02 10 04 CS PS 24 Medical Solutions 6 2 Changes to previous version This page intentionally left blank ...

Reviews: