LITHOSTAR Multiline
RXL2-120.812.01
Page 12 of 24
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
4 - 12
Connections with POLYDOROS SX 65
Connection, Unit - A13 Shock Wave Control Cabinet (blue cable harness)
Type of cable /
connector
Function
In Unit
In Cabinet
ok
Special
Trigger lead
X2,
Power unit
N12.X2
“
Test lead U/2
X5,
Power unit
N12.D91.X2
5)
“
Test lead U
meß
X4,
Power unit
N12.D91.X2
5)
“
Test lead I
meß
X3,
Power unit
N12.D91.X3
5)
EK20 special
Fan power supply
X40, Power unit
N12.K4
EK20 special
Coupling
X12, coupling assembly N12.K5
9 pole, Canon
Coding
X13, Power unit
N12.K103
9 pole, Canon
Shock head sensors
X11, Shock wave head N12.K101
9 pole, Canon
Coding
Shock wave head N12.K102
Special
K26 charger lead
X1,
Power unit
Charger
6)
1 x 4 mm
2
Protective conductor
Power unit
PE, Charger
6)
Type of hose
Water
Cooling system intake Shock wave head
5 Exchanger
Water
Cooling system return Shock wave head
8 Exchanger
Tab. 8
5)
Allow enough cable length so that the boards can be unplugged with the
cables still connected.
6)
- To connect these cables, swing the N12 out towards the front.
2
2
2
2
2
2
2
2
2
1
1
3
3
Fig. 7
- Check that the screw connections of the high
voltage connector K26 are tightened securely
(refer to Fig. 7 on the left)
- Insert the connector of high voltage cable K26 on
the bottom of the charger until it locks audibly into
position.
- The jumper between
and 0V must be present.