Connections/harness locations in the generator cabinet (Part A)
4
•
Route the harnesses in the generator cabinet as shown in Fig. 3 above, and connect
them as indicated in Tables 1 and 2:
Multiline power line connections
•
Connect the power cables for the Multiline in M16 (
/Fig. 3) (refer also to the Multiline
circuit diagram J1042-10):
Connection to contactor K4
•
Strain-relieve the power harness.(shielding under the clamp)
1)
Connections and cable position in the cabinet A13, refer to the fig. on page 3-11.
L1
→
T1
L2
→
T2
L3
→
T3
N
→
K30.N
PE
→
Bolt on M16
AC Generator - A12 Electronics Cabinet - A13 Shock Wave Cabinet Connections
Type cable/
connection
Function
Generator
AC
A12 Cabinet
A13 Cabinet
ok
4 x 4 mm
2
+
Gnd, shielded
MULTILINE
power supply
M16,
L1, L2, L3, N
M16.X1 ,
N
F7, F8, F9
1 x 6 mm
2
Protective con-
ductor
M16.k30, M16,
9 pole Canon
XCS
D 320.X3
VDIM.D5.X13
9 pole Canon
Act. value
interface
D100.X200
N11.D6.X8
3 x 1,5 mm
2
,
shielded
A13 Power
supply
M16.X2.
L3, N, PE
N12.X1.
L, N, PE
1)
1 x 6 mm
2
Protective con-
ductor
M16, base
1)
Canon
Controller
N11.D11.k104
N12.k104
1)
Tab. 1
System Operating Console - Generator Connections
This cable is located in the “System operating console-Electronics Cabinet” zipper hose
Type of cable/
connector
Function
In Control
Console
in AC Generator
ok
15 pole, Canon
System ON/OFF
Door
D160.X9
Tab. 2
D
B
B
5
D
5
C
7
E
6
5
2
5
2
4
2