![Siemens LITHOSTAR Multiline Installation Instructions Manual Download Page 40](http://html.mh-extra.com/html/siemens/lithostar-multiline/lithostar-multiline_installation-instructions-manual_380652040.webp)
LITHOSTAR Multiline
RXL2-120.812.01
Page 18 of 20
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
3 - 18
Connections with POLYDOROS LX 30
Ultrasound option
For 230 V Sonoline line voltage
- Connect the cables to the power line terminal NK 1, 2, 3 (Fig. 15)
- Shielding see A/Fig. 15
For 115 V Sonoline line voltage
(the required parts are located in the 115 V power supply packet)
- Attach the supplied grounded instrument connector (female connector) and the nozzle
to the 115 power supply cable routed into the case frame by the customer.
- Connect the grounded instrument connector (male connector) and the nozzle to the
power line cable from the corrugated hose.
- Plug the two grounded instrument connectors together (corrugated hose - power
supplies).
•
Connect the protective conductor to the clamp (figure 15).
•
Connect the control cable to D31.X28 (figure 15). This is possible only after moving the
system.
•
Route the fiber optic and video cables to the control cabinet. (Refer to the installation
instructions for fiber optic cables /module RX0-000.038.02...).
•
Route the video cable into the control cabinet (not used).
•
Connect the ultrasound tableside control to SONOLINE.
•
Connect the ultrasound probe and place the probe in its holder.
•
Move board D6 slightly away from N11 to facilitate this.
•
Plug the fiber optic cable into N11 on board D6 (according to color code). Secure the
fiber optic cable with cable clamps to the right side of the cabinet.
•
Connect the corrugated hose to the Sonoline.
Unpack and assemble the SONOLINE according to the instruc-
tions.
Type of Cable Function
Ultrasound
system
Multiline
ok
Unit
Cabinet A12
1 x 4 mm
2
Protective con-
ductor
Rear panel
Protective
conductor
terminal
3 x 0.75 mm
2
shielded
Voltage supply
Rear panel
refer to text
Canon
9 pole
Control cable
Ultrasound
tableside control
D31.X28
Fiber optic
cable
Communication
Rear panel
N11.D6
Video
Rear panel
w/o function
Tab. 11
NOTICE