5 - 1
Siemens AG
RXL2-120.812.01
Page 1 of 2
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Final Work Steps
5
Attaching the remaining panels
5
Installation of the remaining panels is described here as part of the unit installation. How-
ever, to prevent double work, this should be done after start-up has been completed as
described in the Start-up Instructions, RXL2-120.034...
•
Install the two large plastic panel sections on the side of the unit and mount them in place
(
/Fig. 1); to do this, temporarily loosen the two screws on the beveled surfaces of the
rear cover panel.
•
Install the X-ray tube housing and attach them together (
/Fig. 2).
•
Secure the EMC metal shielding with screws to the designated location.
•
Close all front walls of the cabinets.
Fig. 1
Fig. 2