Connections with POLYDOROS LX 30
3 - 13
Siemens AG
RXL2-120.812.01
Page 13 of 20
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Connections/Cable Locations in the Generator Cabinet (Part. B)
3
•
Route the cables in the generator cabinet as shown in Fig. 8 above, and connect them in
accordance with Table 9:
•
Install the yellow cable harness in the generator cabinet below M16. Connect the
shielding of the zipper hose to the
bolt in M16 (
/Fig. 8). To strain-relieve the zipper
hose, attach it with a cable tie to the left intermediate plate of the cabinet near the floor
(
/Fig. 8).
Rotating anode cable connection
•
Route the rotating anode cable (internally required length approx. 1 m) up along the right
side in the frame of the generator cabinet to the anode starter (
/Fig. 8).
•
Connect the braided shielding (external shielding) of the end of the lead to the right side
of the starter unit under one of the clamps (
/Fig. or 1/Fig. 9).
•
Connect the red cable (inner shielding) to (3/Fig. 9).
•
Strain-relieve the cable using cable ties (2/Fig. 9).
•
Connect the rotating anode cable to the breaker of the anode starter (
/Fig. 8):
On breaker, contact:
K31/R2 K31/R4 K32/R2
Anode cable leads:
I / 1
0
II / 2