Connections with POLYDOROS SX 65
4 - 21
Siemens AG
RXL2-120.812.01
Page 21 of 24
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
•
Connection of the radiation indicator lamps (e.g. in front of the door of the
examination room)
If an external radiation indicator is desired for fluoroscopy and exposure, connect the
radiation-dependent indicator lamps (24 V ~, max. 5 W) as follows to board D160:
Connect the radiation indicator lamps for separate rooms
Connect general radiation indicator lamp
for AP1
Connect the temperature monitoring devices and oil pressure switches.
•
Room lighting control
A contactor for the room lighting control can be connected to board D160. To program
the room lighting control, refer to the Start-up instructions, System configuration,
RXL2-120.034...” Site Adjustments” in the section “Room light” as required.
Connect contactor connection to:
Indication with ZB:
Indication with radiation only:
jumper X38.1-2 on D160 plugged in
jumper X38.1-2 on D160 pulled out
for AP1 to:
D160.X61.7
24 V
∼
and D160.X61.8 (0 V
L
)
D160.X64.8
24 V ~ and D160.X64.7 (0 V
L
).
If a temperature switch is present, connect this in series with the
oil pressure switch.
D160.X64.5 (24 V ~) and D160.X64.6 (0 V
L
).