4 - 1
Siemens AG
RXL2-120.812.01
Page 1 of 24
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Connections with POLYDOROS SX 65
4
General information
4
The harnesses are enclosed in zipper hoses and are color coded. The specific target con-
nection points for the harnesses have the same color coding in the cabinets:
- The conductor and cable shields have to be connected at the ends with the
corresponding clip.
•
Do not shorten the harness.
•
Whenever possible, avoid overlapping or looping harnesses in the system and
particularly in the cabinets. Do not store excess cable lengths in the cabinets (EMC)
reasons).
•
Be careful that the cable harnesses do not get caught in the swing-out frames.
•
Water hoses 5 and 8 may only be shortened by max. 5 m.
X-ray generator, AC:
yellow
Shock wave cabinet, A13:
blue
Electronics cabinet, A12:
green
Do not open the connector housing on the cable between the
camera and the VIDEOMED DI-M (D8.X1) under any circum-
stances.
Do not step on the hoses. Avoid small loops and kinks as this
may damage them.
Fig. 1
3rd Monitor
Option
Shock wave
cabinet
green
Cables in zipper hose
Cables fixed at the object
Control console
X-ray
generator
Electronics
cabinet
AC
A13
A12
Green
Unit
yellow
Blue
Patient
monitor
Option
Ultra-
sound unit
Option
CAUTION
CAUTION