System Installation
2 - 13
Siemens AG
RXL2-120.812.01
Page 13 of 14
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Installing the Collimator
2
•
Unpack the collimator.
•
Remove the screws from the transport brackets at the side of the blades close to the
focus.
•
Check the adjustment of the blades close to the focus with the template.
•
Attach the collimator to the tube unit.
•
Plug in connector X1 on the collimator.
•
Secure the wire to the panel using strain relief clamps (1/Fig. 19).
•
Connect the plug.
Installing the Diamentor (Option)
2
•
Remove the cover from the collimator.
•
Insert the measuring chamber and lead the cable out through the opening (1/Fig. 20).
•
Reattach the cover to the collimator.
•
Plug the two connectors together and fix the cable with the cable ties (2/Fig. 20).
Fig. 19
Fig. 20
2
1
1