Connections with POLYDOROS LX 30
Connections/harness locations in the generator cabinet (Part C)
3
•
Route the harnesses in the generator cabinet as shown in Fig. 13 above, and connect
them in accordance with Table 10:
Generator power line connection
Nominal Line Voltage: 400 V ± 10%.
•
Secure the screw type terminals from the service packet to the conductors of the power
cable. If the power cable diameter is small enough, turn the clamps 180° (both curva-
tures opposite to each other).
•
Insulate the harness terminals that have been installed using heatshrink tubing.
•
Remove the cover over the knife breakers F1, F2 and F3 (
/Fig. 13) at the M16 of the
generator.
•
Connect the power line leads L1, L2, L3, N, PE to the fuses F1, F2, F3
(connection
/Fig. 13) and to terminal strip K20. N and
.
•
Strain-relieve the power cable on the bottom plate of POLYDOROS LX. Do not allow this
to connect to any existing shielding on the power cable.
•
Reinstall the cover over the knife breakers.