Connections with POLYDOROS SX 65
4 - 11
Siemens AG
RXL2-120.812.01
Page 11 of 24
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Fig. 6
Connections/cable locations in the A13
shock wave cabinet
4
(blue cable harness)
•
Attach the blue cable harness in the shock wave
cabinet A13 on the front and rear cabinet rack at pos.
6 and 7. with cable clips. Refer to Fig. 6. Attach the
braided shield of the zipper hose on the plate
provided at pos. 8/9.
•
Route the cables for the A13 shock wave cabinet as
shown in the Fig. left and connect them as shown in
Table 8.
•
Secure the cables leading to N12 with cable ties on
the right side of position 50/52.
Make sure the conductors to D19 and D91 are
sufficiently long.
•
For connecting the cables to the charger, you have to
tilt M12 toward the front. Use cable ties to secure the
cables on the right side of position 67/68.
•
Connect the hoses for the water connections at the
input/output point
in the shock wave cabinet per
their designations or as shown in connection Table 8
and secure them with hose clamps.
•
Make sure there are no kinks in the hoses and that
they are secured.
The hoses must not be shortened
by more than 5 m, if required
3
NOTICE
HPE
″
Charger
N12
Shock Wave
Controller
Cooling Unit
Water
Preparation
1
A13
2
3
Connection point
for blue cable har-