LITHOSTAR Multiline
RXL2-120.812.01
Page 10 of 24
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
4 - 10
Connections with POLYDOROS SX 65
•
Check the switch positions on the back of the monitors and ensure that the setting is as
follows:
- Video signal to 0.3 V.
- If no third monitor is present, set both monitors to 75 Ohms.
- If a third monitor is present, set the output of the first monitor on high.
•
Reattach the EMC shielding plate on the N11 D11. Make sure that all the cables are in
the recesses and will not get damaged when reattaching the shielding plate.
System Control Console - Electronics Cabinet Connections (right Cable Harness, green)
Type of cable /
connection
Function
In Control Console
In A12 Cabinet
ok
25 pole, Canon
Controller
Slide-out shelf
N11/D11.K11
9 pole, Canon
4 channel
ECG - monitor
4 channel
ECG - monitor
N11/D11.K103A
Coax video cable
Video input
Monitor 1
in
1st Video moni-
tor
Monitor 1
in
MEMOSKOP,X11
8)
Coax video cable
Video input
Monitor 2
in
2nd Video moni-
tor
Monitor 2
in
MEMOSKOP,X12
8)
Coax video cable
Video connection
Monitor 1
⇒
Monitor 3
3rd Video moni-
tor
Monitor 1
out
Monitor 3
in
9)
7)
o to 3rd
monitor if
configured
10)
Coax video cable
Hardcopy input
MULTISPOT
9)
MEMOSKOP, X10
8)
9 pole, Canon
Hardcopy control
MULTISPOT
9)
MEMOSKOP, X4
8)
Tab. 7
connect to the monitor, only if a third monitor is present.
7)
Options
9)
The coaxial cable should be attached with a cable clip to the rail in the electronics
cabinet A12. Electrically isolate the coaxial plug from the rack of the cabinet.
10)
•
After connecting the MEMOSKOP cabling, reinstall the MEMOSKOP cover panel.
The cable shielding must be under the corresponding clamp.
8)
4
9
10
10
10
10
10