![Siemens LITHOSTAR Multiline Installation Instructions Manual Download Page 18](http://html.mh-extra.com/html/siemens/lithostar-multiline/lithostar-multiline_installation-instructions-manual_380652018.webp)
LITHOSTAR Multiline
RXL2-120.812.01
Page 10 of 14
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
2 - 10
System Installation
Back panel of the PC
2
Positioning/Connecting the Operating System Unit (PC)
2
•
Open the door on the console table and pull out the PC slide-out shelf.
•
Position the operating system unit on PC slide-out shelf and secure it from below.
•
Connect the cables for the mouse, the keyboard, the monitors and the DUEP cable at the
PC (Fig. 17).
•
Connect the power line cable to the operating system unit.
•
Insert the slide-out shelf. Pay attention to the cable routing in the console.
Cover the cable outlet
2
•
In areas where UL regulations apply,(e.g. USA) you must cover the cable outlet with the
extra aluminum sheet located in the operating console as fire protection.
Positioning/Connecting DICOM Connect (Option)
2
•
Perform installation and startup according to the manual RXL2-120.814.02..
Fig. 17
DICOM can be built into the LITHOSTAR Multiline only if no
MULTISPOT is installed.
LM - Control Unit
X51
6
1
2
4
7
X50
1 Keyboard port
2 Mouse port
4 Monitor power line connector
6 Monitor port
7 Power line connector
X50 Respirator-ECG signal interface
X51 DUEP port
NOTICE