System Installation
2 - 11
Siemens AG
RXL2-120.812.01
Page 11 of 14
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Positioning/Connecting the Multiformat Camera (Option)
2
•
Connect the Multispot per the designations and place it on the console table as shown in
Fig.15 and tighten the screws. The cables can be found in the console table (power,
video, control cable).
Positioning/Connecting the Diamentors (Option)
2
•
Plug the power cable into the multiple socket outlet of the console desk (slide-out-shelf).
•
Lead the cable up through the free opening.
•
Connect the cable routed from the measuring chamber of the Diamentor via the adapter
cable supplied.
•
Connect the cable routed from power supply unit at the Diamentor.
•
Position the Diamentor on the console desk.
The two Allen screws in the base of the Multispot have to extend
into the drilled holes of the intermediate plate in the operating
cabinet. Depending on the customer's request, the drilled holes
on the front or on the back can be used.
NOTICE