Connections with POLYDOROS LX 30
3 - 15
Siemens AG
RXL2-120.812.01
Page 15 of 20
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
•
Slide the H1 high voltage transformer into the generator cabinet.
•
Change the position of the transformer mount (Fig. 10), lift it, and lock in the bayonet
connectors (1/Fig. 10) of the mount.
•
Connect the high voltage cable shielding to (Z/Fig. 8) (if the shielding is not accessible in
the clamping area, remove the outer insulation).
The clamps are located in the generator service packet.
Fig. 11
•
Use cable ties to strain-relieve the cables
(3/Fig. 11).
•
Carefully remove the plastic guard plate for board
D220 with
symbols (1/Fig. 11).
•
Connect the protective conductor strap to the
H1.
(2/Fig.
11).
•
Attach the (primary) leads H1.U1 and H1.V1 of the
inverter module per the designation onto H1
(3/Fig. 12).
•
Route the cable connected to D220.X41
(4/Fig. 11/12) down to the D160 board and connect
it to X44.
•
Clamp the braided shield of the cable under the
clamp (5/Fig. 11).
•
Use cable ties to strain-relieve the cable at H1
(6/Fig. 12).
Fig. 12
•
Plug the flat cable X1 from board D100 to
connector X1 on board D220 (7/Fig. 12) and strain-
relieve it on the cover (shielding underneath the
strain relief).
•
Reinstall the guard with the
symbols
(1/Fig. 11).
•
Check the nuts on the protective conductor studs
(8/Fig. 12) and tighten them, if necessary.