LITHOSTAR Multiline
RXL2-120.812.01
Page 6 of 20
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
3 - 6
Connections with POLYDOROS LX 30
Fig. 4
Connections of the options and their
cable locations in the A12
electronics cabinet and generator AC
3
•
Attach the special cables for the 3rd monitor to the
rail at the bottom of the electronics cabinet A12.
Install the cables as shown in Fig. 4 and connect
them according to table 4.
Patient monitor connection
•
Position the patient monitor in the location provided
per the project planning guide (e.g. ceiling support).
•
Plug the cable into the "SYNC" connector of the
patient monitor.
•
Connect the patient monitor to its own power supply,
i.e. one that is independent of the LITHOSTAR
Multiline.
Do not establish any protective
conductor connections to the
LITHOSTAR.
9
CAUTION
VIDEOMED
-DIM-
Rail for cable harness
Cable harness, green, from the operating console
Cable harness, green, from the unit (K3, K4....)
Cable harness, green, from the unit (K8, K9....)
3rd monitor on the monitor trolley (option)
Video cable 3rd monitor
Power Supply
M16
System
Controller
N11
Image Storage
8
7
6
9
4
5
10
A12