- 46 -
14.2. External Adjustment
External adjustment is carried out using an external mass standard of the right accuracy and
weight value, which value depends on balance model and capacity. Correction is carried out
semi-automatically, successive process stages are signalled with prompts.
Caution: External adjustment is possible for balances that are not a subject to the
conformity assessment (verification).
Process stages:
•
Enter <Adjustment> submenu, next select “External adjustment” option.
•
Message <Remove weight> is displayed.
•
Unload the weighing pan and press
key. The balance determines start mass,
message
<Adjustment; Please wait…>
is displayed.
•
Upon completed start mass determination, message <Put weight> is displayed along
with particular value of mass standard assigned to the balance.
•
Load the weighing pan with the required adjustment weight and press
key.
•
Upon completed procedure, message <Remove weight> is displayed.
•
Take the weight off the weighing pan, the balance displays the <Adjustment>
submenu.
14.3. Operator's Adjustment
Operator's adjustment can be carried out using any adjustment weight, mass of which
ranges between 0.3 Max and Max. Operator's adjustment and external adjustment
procedures are likewise with one exception, before operator's adjustment start, an edit box
for entering mass of the used adjustment weight is opened.
Caution: Operator's adjustment is possible for balances that are not a subject to the
conformity assessment (verification).
In order to start operator's adjustment, enter <Adjustment> submenu and select <User
adjustment> option. Then follow the commands displayed on a screen.
14.4. Adjustment Test
<Adjustment test> function enables comparing the result of an internal automatic adjustment
with the value recorded in balance factory parameters. The comparison is used for
determining balance sensitivity drift over time.
14.5. Automatic Adjustment
Enter this menu to specify conditions initiating an automatic adjustment. Options:
None – automatic adjustment inactive.
Time – adjustment takes place in time intervals declared in menu <Automatic
adjustment time> (14.6).
Temperature – adjustment is triggered by temperature change.
Both – adjustment is triggered by both, temperature changes and time.
Caution: This parameter settings can be modified only for balances that are not a subject to
the conformity assessment (verification).