- 106 -
32.
32. AMBIENT CONDITIONS
Parameters group which has been designed to enable you to turn on readout of ambient
conditions recorded by the ambient conditions module. Using this submenu, you can also
specify tolerance high and low threshold of both temperature and humidity, and a change
rate for these values per hour. The entered values are referred to indicted ones. Next on the
basis of values comparison respective pictograms are displayed informing you whether the
sensor-read values are comprised within permissible limits or not.
Particular sensors settings:
−
Ambient conditions recording interval [min]: parameter enabling you to specify record
frequency of sensor-registered data, and to determine how often the displayed
pictograms (providing information on ambient conditions state) are to be refreshed, by
default 30 min value is set.
−
Internal sensor 1: enter this parameter to specify tolerance temperature values and
balance temperature change rate.
Min temperature
– min balance temperature, for lower
temperature values the thermometer pictogram is red.
Max temperature
– max balance temperature, for higher
temperature values the thermometer pictogram is red.
Temperature delta t/h
– maximum balance temperature change
rate, for higher change rate the balance displays blinking red
thermometer pictogram.
CAUTION: In order to enable/disable parameter visibility go to service menu.
−
Ambient conditions module: parameters group which has been designed to enable you
to turn on readout of data recorded by a balance-connected ambient conditions
module, and to specify tolerance values for this module.
Ambient conditions module
- parameter specifying module’s
working mode. Available options: N
one
– readout of ambient
condition module's indications turned off;
Record
– readout and
record of indications into module's database turned on;
Record
and alert
– readout and record of indications into module's
database turned on, option of display of warnings informing on
ambient conditions change turned on (the warnings are
displayed on the basis of tolerance high and low thresholds).
Additional temperature sensor
– turning on readout of
indication of the additional temperature sensor connected to the
ambient conditions module.
Min temperature
– minimum temperature value, for lower