- 53 -
A computer keyboard of USB type can be connected to a balance, this enables easier and
quicker editing of the printouts. Willing to insert any text, it is necessary to select a respective
menu option and, using the keyboard, type the text.
15.4. Variables
Variable is defined as alphanumeric data which can be linked to printouts, products or other
information related to weighing. Every variable is characterized by its content, the content
must be inserted. Variables are used for entering various data during the weighing process,
e.g. serial number or batch number. The program allows to enter 3 variables. Each can
consist of 31 characters maximum.
In order to insert a variable content, you need to enter variable settings (VARIABLE 1,
VARIABLE 2, VARIABLE 3) and insert the respective values using arrow keys of the balance
keypad or a computer keyboard. Procedure for entering texts is the same as for non-
standard printouts.
16.
WORKING MODES – General Information
PM series balances feature the following working modes:
Weighing
Means of operation: weight of a load is determined through an indirect
measurement. The balance measures gravitational force which attracts the load.
The obtained result is processed and displayed on the balance screen in a
digital format.
Parts Counting
Means of operation: based on a weight value of a single piece it is possible to
count pieces quantity; necessary conditions: the weight value of a single piece is
determined with sufficient accuracy, the mass of all the pieces is the same.
Checkweighing
Means of operation: control of sample's mass carried out with reference to
specified thresholds, it is necessary to give the value of min threshold <LO> and
max threshold <HI>.
Dosing
Means of operation: it is necessary to specify sample’s target mass to be
obtained by dispensing (pouring, filling, etc.).
Percent weighing
Means of operation: control of percent ratio of a test sample in relation to a
reference sample. Obtained data provides info on how much the test sample
differs from the adopted reference sample.
Solids density
Means of operation: based on Archimedes principle, a balance determines
density of solids. The mode requires use of an optional density determination kit.
Liquid density
Means of operation: based on Archimedes principle, a balance determines
density of liquids. The mode requires use of an optional density determination
kit.
Animal Weighing
Means of operation: mass measurement carried out with use of filters which
dampen animal moves on a weighing pan, the filters enable you to obtain a
correct measurement result.
Statistics
Means of operation: carried out measurements are used to calculate statistical
data such as Min, Max, deviation, etc.
Summary of Contents for PM 10.C32
Page 7: ...7 1 GENERAL INFORMATION 1 1 Dimensions PM xx C32 d 0 01g PM xx C32 d 0 1g...
Page 8: ...8 PM xx C32 d 0 5g d 1g Indicator...
Page 125: ...125...
Page 128: ...128...
Page 129: ...129...