- 56 -
o
YES, weight value of single piece to be updated,
o
NO, weight value of single piece not to be updated.
Means of operation of ACAI function:
1.
pieces quantity upon adding must increase,
2.
pieces quantity upon adding cannot increase more than twice,
3.
pieces quantity upon adding must be comprised within
±
0.3 tolerance of total value,
4.
the weighing result must be stable.
For information on how to operate the remaining settings and parameters read section 11
of this user manual, i.e.
WEIGHING Mode Settings
.
17.2. Parts Counting – Quick Access Keys
Each working mode features automatically displayed set of default keys. The set can be
modified by assigning out-of-list quick access keys to the on-screen buttons. Such operation
requires particular permissions level.
List of quick access keys it to be found in section 8 of this user manual.
17.3. Setting Reference Sample Mass by Entering Mass of a Single Part
Procedure:
Press
button (Set part mass), <Part mass> edit box featuring an on-screen numeric
keyboard is displayed.
•
Enter respective value and press
key for confirmation, <Parts counting> working
mode is run with automatically set mass of a single part.
Caution:
If the value of entered single part mass is lower than 0.1 of the reading unit, then the
following message is displayed: <Value too low>.
17.4. Setting Reference Sample Mass by Determining Mass of a Single Part