- 33 -
Procedure 1
- upon taking the tared load off the weighing pan:
•
Press
key.
•
the <
Net
> marker gets deleted, new zero point is determined.
Procedure 2
- with tared load resting on the weighing pan:
•
Press
key.
•
the <
Net
> marker gets deleted, new zero point is determined.
•
When the zeroed value is above 2% of the maximum capacity, then the software
displays a respective error message.
Selecting tare value out of PACKAGING DATABASE
Procedure:
•
Press
key located on the operation panel.
•
Wait for a respective window to open, next press
button (PACKAGING).
•
Window with a list of tare values recorded into packaging database opens.
•
Select the packaging that is to be used.
•
Balance returns to the weighing mode. The selected tare value (taken form the
selected packaging record) with '
-
' sign is displayed. <Net> marker appears in the
upper-left corner of the screen.
or
•
Press
button (if provided on the display).
•
Window with a list of tare values recorded into packaging database opens.
•
Select the packaging that is to be used.
•
Balance returns to the weighing mode. The selected tare value (taken form the
selected packaging record) with '
-
' sign is displayed. <Net> marker appears in the
upper-left corner of the screen.
AUTOTARE
Autotare function provides automatic taring of the packaging during the weighing process
when packaging mass for each of the product is different. For detailed description of
Autotare function operation, read later section of this user manual.
Deleting tare
The entered tare value can be deleted by pressing
key located on the operation
panel, or by entering value of 0.00g (read the above description).
11.10. Weighing Profile
In order to make operation with the balance easier, 4 various profiles have been designed.
Parameters for these profiles have been set (and saved) in a way ensuring that the
weighings are carried out optimally for particular requirements and conditions.
Profile parameters concern settings of a given working mode, they are grouped in
parameter:
Setup/Working modes/Weighing/Readout
.
For detailed description of Profile settings read the next section of this user manual.
Profile types:
•
User
– basic profile, filters settings provide quite fast and precise measurement.
Summary of Contents for PM 10.C32