- 14 -
Avoid using cleansers containing any corrosive chemicals, e.g. bleach (including chlorine).
Do not use cleansers containing abrasive substances. Always remove the dirt using
microfibre cloth to avoid damage of protective coating.
In case of a daily maintenance:
1. Remove the dirt using cloth dipped in warm water.
2. For best results, add a little dishwashing detergent.
Cleaning Powder-Coated Components
For preliminary cleaning process stage you need running water or wet sponge featuring
large holes, this will help you to remove loose, heavy dirt.
Do not use cleansers containing abrasive substances.
Next, using cloth and cleanser-water solution (soap, dishwashing liquid) gently rub the
cleaned surface.
Avoid using cleanser without water since it may result with damage of the cleaned surface,
please mind that large amount of water mixed with cleanser is a must.
Cleaning Aluminium Components
While cleaning aluminium components use products acid by nature, e.g. spirit vinegar,
lemon. Do not use cleansers containing abrasive substances. Avoid using hard brush, this
may cause scratches. It is recommended to use microfibre cloth.
While polishing the surface use circular movements. Use clean, dry cloth to make the
surface shine.
3.6. Connecting the Weighing Instrument to the Mains
The weighing instrument can be connected to the mains only with a power supply that
comes standard with a particular model. Nominal voltage of the power supply
(specified on the power supply data plate) has to be compatible with the power from
the mains.
Plug the balance to the mains, i.e. connect the power supply to the socket first, next connect
its connector to a port located on the left side of the indicator.
After connecting the balance to the mains, RADWAG logo is displayed. Next you shall see
the home screen and ZERO indication with a respective reading unit (the reading unit
depends on the balance model). During the balance start, the test of an internal adjustment
mechanism takes place (the internal adjustment weight is put down and lifted up). If the
indication is different than zero, press
key.
Caution: In case of verified balance, the automatic adjustment/calibration occurs at start
right after you press the on/off key.
3.7. Temperature Stabilization Time
Before start of the measuring processes, it is necessary to wait until the balance reaches
thermal stabilisation.
Balances stored in much lower temperatures, than the workroom temperature, before being
plugged to the mains (e.g. during winter period) must be subjected to thermal stabilisation.
The thermal stabilisation period takes about 8 hours. During the thermal stabilization, the
indications on the screen can change. Potential workroom temperature change shall occur
gradually and slowly in the course of the weighing instrument operation.