- 34 -
•
Fast
– profile set to enable fast measurement of mass of any size regardless of the
selected working mode. Fast profile gets activated automatically at the first balance
start-up. For Fast profile, the parameters settings enable obtaining the final result as
fast as possible.
•
Fast dosing
– profile intended for dosing, it is set to enable fast dispensing of product.
•
Precision
– profile set to enable precise measurement of mass of any size regardless
of the selected working mode. For Precision profile the weighing process takes a lot of
time, but this guarantees the most precise weighing result.
Caution: as an operator you can modify all profile settings only for User profile, when it
comes to the remaining default profiles (Fast, Fast dosing and Precision) they can only be
modified partially.
At the top of the screen a respective pictogram (letter) is displayed, it informs which profile is
currently operated. The profile can be selected separately for a given working mode. The
most recently used profile is remembered for each working mode (along with all introduced
modifications), the mode, when activated, is run with the most recently used profile.
Profiles pictograms:
No. Pictogram Description
1
U
User
profile
2
F
Fast
profile
3
D
Fast dosing
profile
4
P
Precision
profile
Profile selection procedure:
o
Press
key.
o
List of profiles is displayed.
o
Select demanded profile (grey background), to do it press either
or
button.
o
Press
key to confirm
.
Summary of Contents for PM 10.C32