- 113 -
Set Min Checkweighing Threshold
Format:
DH_XXXXX CR LF
, where:
_
- space
,
XXXXX
- mass format
Response options:
DH_OK CR LF - command carried out
ES CR LF
- command not recognised (mass format incorrect)
Set Max Checkweighing Threshold
Format:
UH_XXXXX CR LF
, where:
_
-
space
,
XXXXX
- mass format
Response options:
UH_OK CR LF - command carried out
ES CR LF
- command not recognised (mass format incorrect)
Give Value of Min Checkweighing Threshold
Format:
ODH CR LF
Response:
DH_MASS CR LF
- command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
D
H
space
mass
space
unit
space
CR
LF
Mass
- 9 characters, right justification
Unit
- 3 characters, left justification
Give Value of Max Checkweighing Threshold
Format:
OUH CR LF
Response:
UH_MASS CR LF
- command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
U
H
space
mass
space
unit
space
CR
LF
Mass
- 9 characters, right justification
Unit
- 3 characters, left justification
Set Mass Value of a Single Item (for PARTS COUNTING mode only)
Format:
SM_XXXXX CR LF
, where:
_ -
space
,
XXXXX
- mass format
Response options:
SM_OK CR LF - command carried out
SM_I CR LF
command understood but not accessible at this moment (e.g. other than
PARTS COUNTING mode operated)
ES CR LF
- command not recognised (mass format incorrect)
Set Target Mass Value (e.g. DOSING mode)
Format:
TV_XXXXX CR LF
, where:
_
-
space
,
XXXXX
- mass format
Response options:
TV_OK CR LF - command carried out
TV_I CR LF
- command understood but not accessible at this moment (e.g. mode other
than DOSING)
ES CR LF
- command not recognised (mass format incorrect)
Set Reference Mass Value (e.g. for PERCENT WEIGHING)