- 43 -
Upon procedure completion, the parameter screen is displayed, value informing on recorded
autotest number (to be found in <RESULT> field) is increased by <1>.
To view particular results, press <RESULTS> field and select the given autotest.
To print a report go to <Details> screen and press printer pictogram that is to be found in the
bottom.
Report example:
...............................................................
----------- Autotest GLP: Report -------------
Balance type xx
Balance S/N 123456
Operator Admin
Software version Lx.x.x
Date 2017.03.10
Time 09:17:16
------------------------------------------------------
Measurements quantity 10
Reading unit 0.01 g
Adjustment weight mass 1402.094 g
Filter Average
Value release Fast + reliable
------------------------------------------------------
Deviation for max -0.018 g
Repeatability 0.0028 g
Signature
........................................