- 118 -
Format:
PC CR LF
Command:
PC CR LF
- send all implemented commands
Response:
PC_A_”Z,T,S,SI…”
- command carried out, the indicator displays
all implemented commands.
Give Balance Type
Format:
BN <CR><LF>
Response options:
BN_A_”x” <CR><LF>
- command understood, response: weighing device type
BN_I <CR><LF>
- command understood but not accessible at this moment
x
- weighing device type (inserted in between inverted commas), with general balance type
in front
Example:
Command:
BN <CR><LF>
- return balance type
Response:
BN_A_”PM C32”
- balance type - ”PM C32”
Give Max Capacity
Format:
FS <CR><LF>
Response options:
FS_A_”x” <CR><LF>
- command understood, response: Max capacity
FS_I <CR><LF>
- command understood but not accessible at this moment
x
- Max value of reading units (in between inverted commas)
Example:
Command:
FS <CR><LF>
- return Max capacity
Response:
FS_A_”10000.00”
- Max capacity: ”10000 g”
Give Program Version
Format:
RV <CR><LF>
Response options:
RV_A_”x” <CR><LF>
- command understood, response: program version
RV_I <CR><LF>
- command understood but not accessible at this moment
x
- program version (in between inverted commas)
Example:
Command:
RV <CR><LF>
- return program version
Response:
RV_A_” 1.1.1”
- program version: ”1.1.1”
Set AUTOZERO Function
Format:
A_n <CR><LF>
Response options:
A_OK <CR><LF>
- command carried out
A_E <CR><LF>
- error in-course of command execution, no parameter or incorrect
format
A_I <CR><LF>
- command understood but not accessible at this moment
N
- parameter, decimal value determining autozero settings:
n
→
0 – autozero off
1 – autozero on
Caution: Command changes settings for a current working mode.
Example:
Summary of Contents for PM 10.C32
Page 7: ...7 1 GENERAL INFORMATION 1 1 Dimensions PM xx C32 d 0 01g PM xx C32 d 0 1g...
Page 8: ...8 PM xx C32 d 0 5g d 1g Indicator...
Page 125: ...125...
Page 128: ...128...