— A-6 —
Appendix 3: Initializing Driver Unit
l
When troubleshooting or replacing Motor or Driver Unit, initializing Driver Unit may be
necessary.
l
When initializing Driver Unit, follow procedures described hereafter.
l
Use Handy Terminal FHT11 for inputting command.
l
Procedures
Figure A-7
(2) Initialize the Driver Unit with “SI” command.
(1) Note down parameter settings and channel programs.
(3) Input the parameters and programms again.
Explanations
1 Read out parameter settings and channel programs and note down them. Especially
“PA” value is important.
(1)
Connect the Handy Terminal FHT11 to CN1 connector of the Driver Unit and turn on the power.
¯
(2)
Monitor the parameters with “TS0” command.
¯
(3)
After monitoring, turn the power off.