Software-settings
•
Open the menu:
Extras/Analog output configuration
(1).
•
The overview screen appears. This menu shows, how much analog-outputs
are available and how the analog outputs are occupied.
At the overview screen all analog-outputs are listed. The amount of the
analog-outputs is dependent from the amount of the installed IO-modules.
Every IO module has 4 analog-outputs. If two IO modules are installed, the
entire amount of analog-outputs is eight. At the list, the user finds the
information which analog signal is carry out at which analog-output
channel. The follow list-notation can be found (example):
Reference:
1.
The number of the IO module. 1 means, first IO module.
2.
The value of the analog-output. 3 means, the third analog-output.
3.
The sample point, from where the signal comes. SP1 means, the first sample
point.
4.
The signal-name. CH4 means, that this analog-output transfers CH4 signal.
This example means: At the third channel of the first IO module, the CH4 signal
from the first sample point will be transferred.
At the sketch below, the menu-screen and the analog-outputs at the first, of two IO
modules are shown.
1
2
3
7
4
AO1
AO2
AO3
AO4
I/O 1/3 SP1 / CH4
1
2
3
4
AO1
AO2
AO3
AO4
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...