Dilutable gases and the actual values (2):
This show the
dilutable gases and the calculated undiluted actual gas
concentration value (C
pure
).
d.
To fine calibrate the dilution-unit select the dilutable gas and
impinge the analyser with the calibration gas. Wait until the gas
concentration will not change anymore. Shift with the right/left
arrow keys the dilution factor, until the actual value has the gas
concentration of the calibration gas cylinder.
Software-settings
The operator can do following settings.
•
Threshold: The operator can change the threshold. If the value is above this
threshold, the dilution will turn on.
o
The path for this setting is:
E
XTRAS
/M
EASUREMENT CONFIGURATION
.
o
Marked the setting
arrow keys the value can be changed.
o
Leaving the menu and confirm the modification.
Setting the dilution cycle
In the default setting the dilution is set automatically. This means, the first cycle will
be measured dilute. If the gas concentration is under the threshold, the next cycle
will be measured undiluted. If the sample gas concentration will reach the
threshold again, the dilution will be activated. During the activation, the
measurement value will be hold for 90 sec. At the display, the dilute measurement
-values will turn on
again.
90 sec.
Threshold
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...