Plug connector definition:
WARNING
Electric voltage
Power the system down and protect from reconnecting,
before starting maintenance work.
Slit screws
Stripping length:
7 mm
Tightening torque min.-max.:
0,5-0,6 Nm
Conductor cross sections, which can be used:
Type of electric line
Conductor cross section min.-
max.
Solid
0,2-2,5 mm² (30-12 AWG)
Stranded
0,2-2,5 mm² (30-12 AWG)
Solid with ferrule (with/ or without
plastic)
0,25-2,5 mm²
Information for cables, which go through the cable gland M16:
It is recommended to use only electric lines with ferrules.