(*2): Not only status numbers 4 to 15, but all status numbers larger than the number
of installed sample points will start the
-
sample points, then statu
-
-
will purge the sensors, then it will close all solenoid valves and switch off the gas
pump. When the status number changes back to a value less or equal to the
afterwards the selected sample point will be measured.
-
ily be used to initiate just a zeroing without any
-
and without changing the sample point.
Example: - status number=1 (for any time period, recommended max. 1 hour)
- status number=15 (for a few seconds, recommended min. 10 seconds)
- status number=1 (for any time period, recommended max. 1 hour)
After installation and power-up of the analyzer few steps should be processed in
order to operate the instrument properly.