Connection of the external control via 4-20 mA input signals
The signal inputs built a 4-bit binary number: I4 I3 I2 I1: 0-11 mA=open=0;
11/12-20 mA=closed=1.
Status of
external signal
source
Status number
Description
I4 I3 I2
I1
-
-
0
0
0
0
0
Automatic sampling point switching
0
0
0
1
1
Analyzer is sampling the point SP1 (*1, *2)
0
0
1
0
2
Analyzer is sampling the point SP2 (*1, *2)
0
0
1
1
3
Analyzer is sampling the point SP3 (*1, *2)
0
1
0
0
4
Analyzer is sampling the point SP4 (*1, *2)
0
1
0
1
5
Analyzer is sampling the point SP5 (*1, *2)
0
1
1
0
6
Analyzer is sampling the point SP6 (*1, *2)
0
1
1
1
7
Analyzer is sampling the point SP7 (*1, *2)
1
0
0
0
8
Analyzer is sampling the point SP8 (*1, *2)
1
0
0
1
9
Analyzer is sampling the point SP9 (*1, *2)
1
0
1
0
10
Analyzer is sampling the point SP10 (*1, *2)
1
0
1
1
11
-
1
1
0
0
12
-
1
1
0
1
13
-
1
1
1
0
14
Remote reset of all system alarms
1
1
1
1
15
Analyzer is
-
-
-
-
-
I4
+
I3
I2
I1
+
+
+
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...