MRU GmbH
SWG 100 User Manual
Page
36
of
141
•
List of the installed analog output ports (1):
At this list the operator can
select, which analog-output should be configured. To select an analog-
output, move the up/down arrow keys.
•
Details (2):
Here the operator can open the configuration screen for the
selected output port. To open the details-menu, push
F1
or
OK
.
•
Auto configuration (3):
If the operator pushes this key (
F2
) the output pins
would be configured by default values.
•
Test (4):
The test-menu is for testing the analog outputs. To open the test-
menu, push
F3
or
OK
.
The navigation inside the menu taken place with the
up/down
arrow keys. With the
OK
key, the analog-output can be edited:
+
The settings for an analog-
menu. See the screenshot below.
o
Section (1):
Shows the operator, which
analog-output will be changed.
o
Sample point (2):
The operator can select
the sample point. This means, that the
analog-output signal from the choose
sample point will be transmitted to the
process control system.
o
Meas. Item (3):
At this point the operator
selects the measurement item, which
should be transmitted. At the chart below,
the typical measurements items are listed.
Basically, all measurement-channels, which
can be measured, can be selected at this
point.
o
Meas. Unit (4):
This point shows, which unit
the transmit signal have. This point cannot
be changed.
o
Minimum (4mA) (5):
Here the operator
entering the equivalent measurement
value, for a current of 4 mA.
o
Maximum (20mA) (5):
Here the operator
entering the equivalent measurement
For navigation
and selection
1
2
3
4
5
6
7
8
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...