Software-side
•
Open the menu:
E
XTRAS
/A
UX INPUT
CONFIGURATION
(1).
•
The overview screen appears. At the overview-screen, the installed AUX-
inputs are listed. Every IO module has four AUX-inputs. At the default
settings all AUX-inputs are deactivated (OFF at the overview-screen).
•
To activate an AUX-input, push
F1
(=details). The different IO-inputs can be
selected with the up/down arrow keys.
•
After one AUX-input is selected, the measurement item must be activated.
To do this, turn the measurement item from OFF to definable.
•
The measureme
configuration, where the user can configure by himself.
contains some pre-configurated settings, like for
temperature, or pressure sensors.
given
at the screenshot below.
•
Section (1):
Show the operator, which analog-input
will be changed.
•
Meas. item (2):
The operator can select the sample
point. This means, that the analog-output signal
from the choose sample point will be transmitted to
the process control system.
•
Item name (3):
At this point the operator selects the
measurement item, which should be transmitted. At
the chart below, the typical measurements items
are counted. Basically, all measurement-channels,
which can be measured, can be selected at this
point.
•
Meas. unit (4):
The dimension of the measurement
•
Sample point (5):
This point shows, which unit the
transmit signal have. This point cannot be changed.
•
Minimum (4mA) (6):
Here the operator entering the
equivalent measurement value, for a current of 4
mA.
•
Maximum (20mA) (6):
Here the operator entering
the equivalent measurement value, for a current of
20 mA. The IO module will create the linear relation.
•
OFF (6):
Leave the menu. Alternate push
F2
.
+
+
1
2
3
4
5
6
7
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...