Mounting and installation of the gas cylinders for the auto calibration function
Mounting and installation of the gas cylinders for the auto calibration function
The auto calibration function allows the calibration of the analyzer. To use the auto calibration
function it is necessary to install the calibration gas bottles on the analyzer.
WARNING
High pressure
•
Only educated staff is allowed to install the gas cylinders on the
analyzer.
•
All gas cylinders must be equipped with a pressure reducer. The
pressure must be set to 500 mbar.
Recommended gas concentration for the gas bottle:
CO
2
~39,95 Vol.%
CH
4
~60,00 Vol.%
H
2
S
~500 ppm
(1)
Primary pressure
:
500mbar
(1)
Cal inlet port (scope of supply)