Position 4: PTFE Filter (#51513)
Required materials: PTFE filter (#51513) contained in the service set.
Required tools: needle-nose pliers.
Exchange interval: Exchange necessary, if gas flow is too low.
Mounting direction: Irrelevant
Steps:
Pull the exhausted filter-unit from the viton
tube and exchange it with a new one. If it
necessary, the tube can be solved from the
tube with a needle-nose pliers.
Position 5: Sintered filters (#65988)
Required materials: Sintered filters (65988), contained in the service-set.
Required tools: Spanner (range:17)
Exchange interval: Every 2 month, dependent from the dust- and pollution load.
Steps:
1.
Remove the gas tube. Be sure, that no
process gas emits in the environment.
The upper part from the nozzle can be
removed with a spanner (range 17).
2.
Inside of the nozzle is a sintered filter
tablet with an O-ring. Remove the
exhausted sintered filter tablet and
the O-ring and exchange both parts
through
a
new
one.
The
compensation parts can be found in
the service-set.
3.
The gas tube can now connect with
the nozzle again.
O-ring (#64798)
Sinterfilter
(#65988)
Copper seal
(#61947)
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...