MRU GmbH
SWG 100 User Manual
Page
63
of
141
•
Phase duration:
Entire Stand-by time (Purging time + Quiet time = Phase time).
•
Purging time:
The time, to purge the analyser with ambient air, through the zero gas inlet.
•
Quiet time:
The time, where the analyser is in the pure stand-by mode.
STAND-BY
Measuring site valves
Valves closed
Zeroing valve
Valve closed
Duration
Phase duration: 2 min to 24h
Purging time: 30 sec. to 1h
Quiet time: calculated
•
Purging (cycle phase details):
The purging is a separate configuration point to purge the analyser
with ambient air through the zero gas inlet. It can be helpful, if the analyser must switch between a
sample point with different sample gas concentrations.
PURGING
Measuring site valves
Valves closed
Zeroing valve
Valve open
Duration
30 sec. to 1 h
Activated/deactivated a phase
The user has the opportunity to deactivate a phase in the measurement configuration cycle. This
could be necessary for example if a sample point is temporarily not in use. The activation and
deactivation of a phase can be done in the cycle phase details of the concerning phase.
Example for the deactivation of a phase
In this example
=
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...