The diagram above shows the necessary installations for the auto calibration. The mixed gas
cylinder is installed on the left side of the analyzer.
Software side: Adjustment for the option auto-calibration:
1.
Open the menu
-
can be found on the path:
E
XTRAS
/
A
DJUSTMENT
MENU
/
S
ETUP
-
AUTO
.-
CALIBRATION
2.
Select the
up/ down
an empty area can be selected.
left/ right
3.
With
OK
the user is able to set the gas concentrations for the different gas compositions.
First select the gas component, then press
OK
. A blue screen will be appeared. With the
arrow keys the gas concentration can be set.
4.
After the gas composition is set, the interval for the calibration can be set, too.
Arrow keys left/ right:
With these keys the gas
composition can be set.
Arrow keys up/ down:
Single positions for
the different gas compositions can be
selected.
+