MRU GmbH
SWG 100 User Manual
Page
135
of
141
Dilution-unit
The dilution-unit inside the analyser dilutes the sample gas flow with a fix ratio. The unit is
preadjusted. The operator must do a fine calibration of the device, to improve the accuracy of the
measurement. For the fine calibration it is indispensable to tube the vent and all sample gas inlets,
before making the actual calibration. The correct order for the fine calibration is shown in the flow
chart.
Correct installation of the SWG-100
Tubing of the vent
Tubing of all sample gas inlets
Installation of the calibration gas cylinder
Fine calibration of the dilution-unit
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...