Connection of the external control via one 4-20 mA input signal
The user can control the analyzer with only the first 4-20mA input (see sketch
below). The different commands will be given by the changing of the current
signal. The offset-signal is 4 mA. Every 1 mA step describes a condition of for the
external control. Overall the analyzer can take 16 different statuses. The first status
is by 5 mA (4 mA+1 mA) the second is by 6 mA (4 mA + 2 mA) and so on until the
20mA signal is reached.
The connection of the one 4-20 mA signal is a two-wire connection.
Status of
external signal
source I [mA]
Status
number
Description
4
0
Automatic sampling point switching
5
1
Analyzer is sampling the point SP1 (*1, *2)
6
2
Analyzer is sampling the point SP2 (*1, *2)
7
3
Analyzer is sampling the point SP3 (*1, *2)
8
4
Analyzer is sampling the point SP4 (*1, *2)
9
5
Analyzer is sampling the point SP5 (*1, *2)
10
6
Analyzer is sampling the point SP6 (*1, *2)
11
7
Analyzer is sampling the point SP7 (*1, *2)
12
8
Analyzer is sampling the point SP8 (*1, *2)
13
9
Analyzer is sampling the point SP9 (*1, *2)
14
10
Analyzer is sampling the point SP10 (*1, *2)
15
11
-
16
12
-
17
13
-
18
14
Remote reset of all system alarms
19
15
-
-
+
I1
Summary of Contents for SWG 100 biogas
Page 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Page 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Page 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Page 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Page 86: ......
Page 92: ......
Page 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Page 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Page 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...