Zeroing (Cycle phase details):
Z
EROING
configured.
ZEROING
Measuring site valves
Valves closed
Zeroing valve
Valve open
Duration
2min to 1 h
Recommendation
5min., in general not to be changed by user as
depending only on analyzer internal setup
Measurement SPX (Cycle phase details):
In the cycle phase deta
measurement time and the suction delay can be configured. Each sample point can be configured
individually. In the cycle phase details the following times can be set:
MEASUREMENT SPx
Measuring site valves
Valve of selected site is open, others closed
Zeroing valve
Valve closed
Duration
Phase duration: 2 min. to 24 h
Suction/response time: 30 sec. to 1h
Pure measurement: calculated
H2S-low: Activated/protect (Optional)
Stand-by (cycle phase details):
In the cycle
-
configured. In the cycle phase details the following times can be set: